AVOX QC

Quality Control QC for the AVOXimeter 1000E serves as a means to verify that the system is functioning properly. There are two types of QC tests: Optical QC Filters and Liquid QC (LQC).

                Optical QC Filters (Yellow and Orange)

  • Run one of each filter daily prior to patient testing.
  • Run if analyzer is brought from hot to cold temperatures or vice versa.
  • Run whenever the performance of the analyzer requires verification.

Optical QC Ranges are as follows:

  • tHb (g/dL) – [Yellow Filter: 7.8 – 8.2] [Orange Filter: 16.7 – 17.3]
  • %O2Hb (O2 Sat) – [Yellow Filter: 93.7- 96.13] [Orange Filter: 37.8 – 40.2]

Liquid QC

  • Control materials compatible with the AVOXimeter 1000E are the RNA Medical CC 527 or QC 253 (Levels 1, 2, and 3). Other controls may give erroneous results.
  • Run weekly to verify proper storage of cuvettes.
  • Run every time there is a new lot or shipment of cuvettes prior to patient testing
  • QC results should fall within the manufacturer’s reference range found listed on each lot’s package insert.

 

Current QC Log Sheets (Liquid QC) Run WEEKLY

Current QC Log Sheets (Optical QC) Run DAILY